Electrodeposition of iridium in graphite tube as permanent chemical modifier in atomic absorption spectrometry Eletrodeposição de irídio em tubo de grafite como modificador químico permanente em espectrometria de absorção atômica

dc.contributor.authorNaozuka J.
dc.contributor.authorOliveira P.V.
dc.contributor.authorPedrotti J.J.
dc.date.accessioned2024-03-13T01:45:59Z
dc.date.available2024-03-13T01:45:59Z
dc.date.issued2003
dc.description.abstractA tubular electrochemical flow-cell for iridium deposition on the inner surface of pyrolytic graphite tube for permanent chemical modification is proposed. A transversal heated graphite tube was used as working electrode, a cylindrical piece of graphite inserted into the graphite tube as auxiliary electrode, and a micro Ag/AgCl(sat) as reference electrode. Indium solution in 1.0 mol L-1 HCl, flowing at 0.55 mL min-1 for 60 min was used to perform the electrochemical modification. The applied potential to the flow-cell was - 0.700 V vs Ag/AgCl. Scanning electron microscopy images were taken for thermal and electrochemical modified graphite surface in order to evaluate the iridium distribution. Selenium hydride trapping was used to verify the performance of the proposed permanent chemical modifier.
dc.description.firstpage934
dc.description.issuenumber6
dc.description.lastpage937
dc.description.volume26
dc.identifier.doi10.1590/s0100-40422003000600026
dc.identifier.issn0100-4042
dc.identifier.urihttps://dspace.mackenzie.br/handle/10899/37984
dc.relation.ispartofQuimica Nova
dc.rightsAcesso Aberto
dc.subject.otherlanguageAtomic absorption spectrometry
dc.subject.otherlanguageIridium electrodeposition
dc.subject.otherlanguagePermanent chemical modifier
dc.titleElectrodeposition of iridium in graphite tube as permanent chemical modifier in atomic absorption spectrometry Eletrodeposição de irídio em tubo de grafite como modificador químico permanente em espectrometria de absorção atômica
dc.typeArtigo
local.scopus.citations3
local.scopus.eid2-s2.0-3242807442
local.scopus.updated2024-05-01
local.scopus.urlhttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=3242807442&origin=inward
Arquivos